User Tools

Site Tools


atmos:citation:instruments:cip:home

Cloud Imaging Probe

  • Description: Cloud Imaging Probe
  • Manufacturer: Droplet Measurement Technologies
  • Serial Number:
  • Measured Particle Size Range: 25 um - 1.55 mm
  • Owner:Center for Interdisciplinary Remotely-Piloted Aircraft Studies (CIRPAS)

Problems and Solutions

Problem

The CIP was showing housekeeping data as well as mean particle diameters and concentrations, but no images.

Cause

Bad connection between the CAPS card and the M300 computer.

Solution

Reset the CAPS card in the M300 computer.

Calibration Constant

From the July 16, 2011 flight

 67600 to 68300 Seconds from Midnight 

Nose Pitot vs IAS

Slope: 2.231 Offset: 0.00
Slope: 1.163 Offset: 42.939

Wing Pitot vs. IAS

Slope: 2.2504 Offset: 0.00
Slope: 1.3284 Offset:37.09

atmos/citation/instruments/cip/home.txt · Last modified: 2020/01/29 17:25 by 127.0.0.1