King Probe

  • Description: CSIRO-KING Liquid Water Content Probe
  • Manafacturer: Particle Measureing Systems Inc.
  • Model: KLWC-5
  • Serial Number: SN-PMI-1058-0704-86 (Purchased fall 2008)
  • 200 um Drops: - 50% LWC (See Walter Strapp paper, “Measurements of the response of hot-wire LWC and TWC probes to large droplet clouds”)

Instrument Requirements

  • Power: 28 VDC 10 Amps [Heaters] 28 VDC 70-120 Watts [Source: Paul LeHardy]
  • Weigth: 1.8 Kg (4 lbs) [Total weight for control Unit, display and sensor] [Source: Paul LeHardy]
  • Size: width - 6 inch. Height - 12 inch. Depth - 12 inch. [Control Unit] [Source: Davie Delene]
  • Hot Wire Size: 1.92mm Measured October 20, 2010 [Source: Aaron Ness]
  • Hot Wire Diameter Size: 1.92mm, Length - 47.30mm Measured August 25, 2011, new element changed out on June 1, 2011 [Source: Aaron Ness]

Calibrations

  • Calibration plots from calibration done on December 30, 2010 from 4550 to 4880 seconds from midnight.

Problem

During April 8, 2011 flight the signal voltage was negative.

Cause

The wire on the A/D board was connected backwards for this flight.

Solution

The A/D wires were switched. The King probe was tested on the ground to valid that there is positive voltage when the probe is turn on. Labels were put King Probe A/D wires.

Problem

During L3COM flight on November 24, 2004 the King Probe wire broke.

Cause

Normal to sometimes break.

Solution

King Probe wire replace after the flight.

Problem

During the TAMDAR Turbulence flight on September 9, 2005 the King Probe wire broke.

Cause

Flight encounted some heavy rain and a little hail. This resulted in breaking the broken wire.

Solution

King Probe wire replace after the September 13 ferry flight to Fairbanks, Alaska. The King probe power was not turn on during the ferry flights.

Problem

King Probe voltage possibly high. Probe does not match water content measured by other probes.

Solution

Varied voltage on a fan blowing on the hot wire of the probe. Dave and Aaron measured the following on 10/25/11 on data file 2011_10_25_19_24_46: 28.0 vdc on fan read 0.940 on King display, and current mesured 2.56 amps. 23.0 vdc, 0.88 displayed, 2.48 amps. 20.0 vdc, 0.84 displayed, 2.42 amps. 15 vdc, 0.75, 2.29 amps. 10.0 vdc, 0.64, 2.11 amps. 5.0 vdc, 0.27 displayed, 1.43 amps. 0.0 vdc, 0.29, 1.47 amps. re-did 28.0 vdc, 0.96 displayed, 2.57 amps.

Problem