Line 80: | Line 80: | ||
{{: | {{: | ||
- | | + | |
+ | |||
+ | =====Problem===== | ||
+ | December 30, 2021 (12/30/2021 - David Delene). Ground Test had approximately 0.08 volts on slave voltage. | ||
+ | |||
+ | |||
+ | =====Solution===== | ||
+ | 12/31/2021. Replace the Sensor Wire with a new one. Across the middle pin (of the 5 pins) on the Sensor Wire should have almost zero resistance (conduction test, not open circuit). | ||
+ | |||
+ | |||
+ | |||
+ | =====Problem===== | ||
+ | High Slave Voltage (01/14/2022 - David Delene). During the January 14, 2022 IMPACTS flight, the King Probe Slave voltage was above 6.0 volts all during the flight, when the voltage should be 5.0 volts. | ||
+ | |||
+ | |||
+ | =====Solution===== | ||
+ | 01/14/2022. The slave negative connected on the analog to digital board was not secure on the screw connections. Put new pin ends on all slave and signal, positive and negative connectors. | ||
+ | |||
+ | |||
+ | =====Problem===== | ||
+ | High Slave Voltages (01/17/2022 - Jennifer Moore). During the January 17, 2022 IMPACTS flight, the King Prove Slave voltage started with 5.21 volts before take off. After take off, and during the rest of flight the Slave voltage was between 6.1V to 6.2V, which is above the usual 5.0 Volts. Additionally, | ||
+ | |||
+ | =====Solution===== | ||
+ | 01/18/2022. The slave negative and signal negative was labeled incorrectly for the analog to digital board wiring. | ||
+ | * Blue and White wire - Slave high | ||
+ | * Orange and White wire - Slave low | ||
+ | * Black and White wire - Ground (Disconnected) | ||
+ | * All White wire - Signal high | ||
+ | * Green and White wire - Signal low | ||
+ | |||
+ | =====Problem===== | ||
+ | Low Slave Voltages (11/11/2022 - Jennifer Moore). After installing King probe on hot-wire boom during NASA IMPACTS integration, |